Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
The test report appears automatically and it is shown as html page (see below).
Figure 49
NOTE
The report can be printed or saved in the computer.
NOTE
Check that the radiation beam can be seen entirely on the image receptor. The radia-
tion beam should not exceed the receptor.
If the radiation is not shown on the image receptor, calibrate the X-ray unit again and take
another exposure.
If the test report shows that the radiation can be seen on the image receptor, you can exit the
Proline Beam Check program by selecting Exit from the File pulldown menu.
E-26 Planmeca Proline XC X-ray unit
PANORAMIC X-RAY BEAM ADJUSTMENT
Date and time according to
the PC
Detailed view of the upper part
of the active area
Panoramic test report
Beam location in the active area
of the sensor
Detailed view of the lower part
of the active area
Technical Manual