Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Place the beam alignment tool to the sensor alignment tool to horizontal position (see figure
below).
Figure 124
Enter the service setting mode. By touching the arrow down field below the list you can see
more features. Select the primary collimator mode from the Service settings list.
Figure 125
E-66 Planmeca Proline XC X-ray unit
CEPHALOSTAT ADJUSTMENT
Beam alignment tool
Technical Manual