Chapter H - PARTS REPLACEMENT & REPAIR
1.5 Lower shelf cover
a)
Drive the head support to open position.
b)
Drive the positioning mechanism to outmost position).
c)
Remove the chin rest adaptor.
Figure 4
d)
Remove the fixing screws under the lower shelf.
e)
Remove the cover by lifting it up.
Figure 5
H-4 Planmeca Proline XC X-ray unit
COVERS