DISTURBANCES IN CASSETTE MOVEMENT

5.5 Abnormal noise during rotation

Trembling from step motor gen-
erates resonance noise which
is characteristic to the mecha-
nism. Cables from the upper
part of the Z-carriage may also
generate
besides these is abnormal.
6 DISTURBANCES IN CASSETTE MOVEMENT

6.1 Cassette carriage does not move

Cassette motor is not running

6.2 Cassette carriage stops suddenly

The motor stops running. If
error code appears, see error
code explanations.
Motor stops/stalls
Motor/drawing wheel glides
Technical Manual
SYMPTOM
noise.
The
noise
SYMPTOM
SYMPTOM
POSSIBLE REASON
Attachment of the driving sur-
face to the groove.
Function of the cable guide.
POSSIBLE REASON
Keyboard out of order
PCB defective
Motor defective or connector
loose
POSSIBLE REASON
Motor or PCB defective or
motor cable loose.
Gear assembly damaged.
One of the connectors loose.
Linear bearing gets stuck.
Cassette carriage gets stuck to
secondary collimator or another
part of the mechanism.
Grease on driving wheel or
driving surface.