Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
The display shown below appears. The number above the arrow fields indicates the collima-
tor number and the collimator is changed with the Next field. Select collimator number 0.
Figure 27
Darken the room sufficiently so that you will be able to see the image of the radiation beam on
the alignment tool (it is fluorescent and will glow when the radiation beam strikes it), but not so
dark that you cannot see the borders of the alignment rectangle.
Protect yourself from radiation and press the exposure button. The beam image will appear
on the alignment tool. Observe the beam from behind the tube head.
CAUTION
Radiation is generated when the exposure button is pressed. Take adequate protection
measures. Keep the exposure time as short as possible.
The beam image should appear within the borders of the rectangle marked on the alignment
tool (if the beam is not correctly aligned see section 2.4 "Radiation beam adjustment, Plan-
meca Proline XC Panoramic X-ray unit" on page E-16, or section 2.5 "Radiation beam adjust-
ment, Planmeca Proline XC Pan/Ceph X-ray unit" on page E-19).
Beam alignment tool
Radiation beam
Figure 28
Planmeca Proline XC pan/ceph X-ray unit: When the collimator number 0 has been
checked, and if necessary adjusted, select the collimator number 1 with the arrow fields.
E-14 Planmeca Proline XC X-ray unit
UP
DOWN
PANORAMIC X-RAY BEAM ADJUSTMENT
Primary slot
Technical Manual