Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Adjusting the magnet holding force on the cephalometric
position
The holding force of the magnet can be adjusted with the screw located in the middle of the
magnet.
Figure 100
Adjusting the movement friction of the sensor head
The sensor head must be detached from the rotating unit before the movement friction can be
adjusted. Detach all the sensor head cables. Unscrew the three M6x12 ISO 7380 attachment
screws (Fig. 101 and Fig. 102).
Sensor head
attachment screws
Figure 101
E-50 Planmeca Proline XC X-ray unit
Magnet adjusting screw
PATIENT POSITIONING MECHANISM
Technical Manual