Planmeca Proline XC Technical Manual page 108

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

Chapter D - ADJUSTMENT & CALIBRATION
Slide the beam alignment tool to the right of the cassette holder.
Figure 87
Select the collimator number 5.
Figure 88
Stand behind the tube head and protect yourself from radiation. Press and hold down the
exposure button.
The image of the radiation beam from slot number 5 will appear on the alignment tool.
D-54 Planmeca Proline XC X-ray unit
Cassette
holder
Beam
alignment
tool
PLANMECA PM 2002 Proline Cephalostat 18x24
4
5
No. 652232
CEPHALOSTAT ADJUSTMENT
Technical Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents