Planmeca Proline XC Technical Manual page 163

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

PATIENT POSITIONING LIGHTS
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
The Frankfort plane light beam must be located between the two black lines on the alignment
tool. It must be horizontal.
Frankfort plane light beam
7
9 7
. 5 0
N O
O Y
C A
M E
A N
P L
Midsagittal light beam
Focal layer light beam
The focal layer light beam must be positioned so that it is on the black reference line on the
side of the ball phantom.
The midsagittal plane light beam must strike the black line on the front of the ball phantom.
If the light beams do not coincide with the black lines or they are not in focus, they should be
adjusted.
Planmeca Proline XC X-ray unit E-53
Technical Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents