Checking The Head Support Position - Planmeca Proline XC Technical Manual

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

CEPHALOSTAT ADJUSTMENT
7.5

Checking the head support position

Technical Manual
Place the beam alignment tool to the sensor alignment tool to horizontal position (see figure
below).
Figure 134
Enter the ceph exposure mode. The rotating unit will move automatically to the correct posi-
tion for taking cephalometric exposures. Select collimator number 6.
Figure 135
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Beam alignment tool
Planmeca Proline XC X-ray unit E-71