AUTOMATIC EXPOSURE CONTROL (AEC) CALIBRATION
Chapter D - ADJUSTMENT & CALIBRATION
Move the cassette carriage to the loading position by touching the Entry field.
Figure 63
In the darkroom open the cassette and place a film inside the cassette. Close the cassette.
Slide the cassette into the cassette holder.
NOTE
The film density adjustment must be performed in normal panoramic mode. Make sure
that the unit is NOT in the AEC mode.
Select a kilovolt value of 68 and a milliampere value of 7.
Drive the X-ray unit to the exposure position by touching the Ready field. The rotating unit will
turn to the start position and the indicator light will come on.
Figure 64
Take an exposure.
Touch the Entry field to drive the cassette to the loading position, remove the cassette and
process the film in the same way you would do in clinical exposure procedure.
Planmeca Proline XC X-ray unit D-37
Technical Manual