Planmeca Proline XC Technical Manual page 38

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

Chapter B - GRAPHIC USER INTERFACE
Downward movement
Select the Motor down adjustment from the Service settings list.
The display shown below appears.
UP and DOWN
arrow fields
The current downward starting speed of the vertical carriage as a percentage of the maximum
speed (00-99) is shown on the display. The speed indication is relative, 00 meaning the mini-
mum and 99 the maximum possible slow adjustment speed. 00 is not stop and 99 is about
half of the maximum speed of the motor.
Touch the right and left arrows to increase or decrease the speed. The new value will appear
on the display.
The speed can be checked by moving the equipment with UP or DOWN arrow fields. Note
that the direction under modification never accelerates to the full speed but continues at the
slow speed to allow the proper verification of it. The opposite direction operates normally.
Touch the OK field to accept the new speed.