X-Ray Units With Two Fixed Sensor Heads: Panoramic Sensor Head Adjustments - Planmeca Proline XC Technical Manual

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

PATIENT POSITIONING MECHANISM
4.4
X-ray units with two fixed sensor heads: panoramic sensor
head adjustments
NOTE
Adjusting the panoramic position of the sensor head
Magnet stopper
locking screw
Technical Manual
This section applies to pan/ceph units with two fixed sensor heads.
The panoramic position of the sensor head can be slightly adjusted by adjusting the position
of the magnet stopper. Loosen the magnet stopper locking screw and rotate the stopper to
the desired position. Tighten the locking screw.
Magnet stopper
Figure 99
The holding force of the magnet can be adjusted with the screw located in the middle of the
magnet (see Fig. 99).
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Magnet adjusting screw
Planmeca Proline XC X-ray unit E-49