CALIBRATING PANORAMIC SENSOR HEAD
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
On the Segmentation display SWITCH OFF the radiation by deselecting all the sectors with
the arrow fields. The sectors will turn dark.
Figure 56
The pmpcal2 window with text Waiting appears.
Figure 57
Touch the Ready field to drive the X-ray unit into the exposure position.
The text Exposure appears.
Figure 58
Take an exposure. After the exposure the message window shown below appears.
Figure 59
SWITCH ON the radiation by touching the arrows of the deselected sectors again.
Planmeca Proline XC X-ray unit E-29