Planmeca Proline XC Technical Manual page 194

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
The pmpcal2 window with text Waiting appears.
Figure 158
Touch the ready field to drive the X-ray unit to the exposure position.
The text Exposure appears.
Figure 159
The calibration tool image should be evenly grey. If it is not, recalibrate the sensor head. If
recalibration does not help, check that the Dimaxis, DIDAPI and drivers you are using are all
from the same CD rom. Check also the cephalostat adjustments by taking an exposure from
the ear posts (see section 7.5 "Checking the head support position" on page E-71).
Editing the calibration image
In case the calibration image contains horizontal stripes, the image can be edited, i.e. the
selected row can be removed by marking it as "bad". You can enlarge the image with the
zoom function if needed. The image can be moved with the Hand-tool.
Press the Marking tool -button located on the right side of the toolbar and then click the row
that you want to mark with the left mouse button. Note that the row number is shown on the
status bar at the bottom of the window.
Figure 160
E-84 Planmeca Proline XC X-ray unit
Zoom in
Fit image to window
Zoom out
CEPHALOSTAT ADJUSTMENT
Hand
Marking tool
Technical Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents