CEPHALOSTAT ADJUSTMENT
NOTE
4
PLANMECA PM 2002 Proline Cephalostat 18x24
Technical Manual
Orientate the cephalometric beam alignment tool so that it is vertical and then place it in the
cephalometric cassette holder. Position it so that it is on the left of the cassette holder.
Cassette
holder
Beam
alignment
tool
Figure 85
Stand behind the tube head and protect yourself from radiation. Press and hold down the
exposure button.
The soft tissue filter will filter a narrow area of the left side of the radiation beam.
The image of the radiation beam from collimator number 4 will appear on the alignment tool.
The radiation beam must appear within the borders of the rectangle marked on the alignment
tool. If it does not refer to section 2.3 "Primary collimator adjustment, Planmeca Proline XC
Pan/Ceph X-ray unit" on page D-12.
5
No. 652232
Beam
Figure 86
4
5
PLANMECA PM 2002 Proline Cephalostat 18x24
No. 652232
Chapter D - ADJUSTMENT & CALIBRATION
Fourth primary
collimator
Planmeca Proline XC X-ray unit D-53