Planmeca Proline XC Technical Manual page 200

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
You can type any kind of additional information (e.g. name of the clinic) on this window. Con-
firm by clicking the OK button.
The test report appears automatically and is shown as html page.
Tes t R eport , Thu Jun 10 09: 07: 21 2004
Figure 174
NOTE
The report can be printed or saved in the computer.
NOTE
Check that the radiation beam can be seen entirely on the image receptor. The radia-
tion beam should not exceed the receptor.
If the radiation is not shown on the image receptor, calibrate the X-ray unit again and take
another exposure.
If the test report shows that the radiation can be seen on the image receptor, you can exit the
Proline Beam Check program by selecting Exit from the File pulldown menu.
E-90 Planmeca Proline XC X-ray unit
Cephalometric Test Report
Detailed view of the upper part
of the active area
Beam location in the active
area of the sensor
Detailed view of the lower part
of the active area
CEPHALOSTAT ADJUSTMENT
Date and time according
to the PC
Technical Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents