Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
The pmpcal2 window with text Waiting appears.
Figure 60
Touch the Ready field to drive the X-ray unit into the exposure position.
The text Exposure appears.
Figure 61
Take an exposure.
The sensor head is now calibrated. After the exposure has been taken the image is shown in
the Dimax2Tool window.
Figure 62
E-30 Planmeca Proline XC X-ray unit
CALIBRATING PANORAMIC SENSOR HEAD
Technical Manual