Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Cephalostat sensor head with quick connector mechanism -
removing the covers
Unscrew the bottom cover plate attachment screw with the 4mm allen key and remove the
cover plate.
Figure 110
Turn the lock disc from above the locking plate and push in the quick connector mechanism
locking knob. Unscrew the attachment screw of the quick connector mechanism cover with
the 4mm allen key.
Figure 111
E-58 Planmeca Proline XC X-ray unit
Cover
Bottom cover plate
Locking plate
CEPHALOSTAT ADJUSTMENT
Technical Manual