Cassette Motor/Gear; Positioning Motors - Planmeca Proline XC Technical Manual

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

Chapter H - PARTS REPLACEMENT & REPAIR
Changing motor to gear
a)
Slack the stop screw of the rubber clutch
b)
Remove the nuts going trough the rubber plate (4 pcs)
c)
Remove the motor.
d)
Detach the rubber clutch from the motor axle.

2.3 Cassette motor/gear

a)
Remove the outer secondary collimator and back cover of the cassette head.
b)
Remove the loading spring of the cassette motor.
c)
Remove the connector.
d)
Remove the attaching screw of the cassette motors attaching shaft.
e)
Remove the motor.
f)
Remove the spring nipple and attaching axle with bearings.

2.4 Positioning motors

a)
Remove the cover of lower shelf.
b)
Remove all the connectors.
c)
Remove the handles.
d)
Remove the attaching bolts of the positioning mechanism.
e)
Remove the attaching mechanism.
Patient positioning mechanism motor
a)
Remove the panduit beside the patient positioning mechanism motor
b)
Remove the patient positioning mechanism motor with rubber plate.
c)
Remove the rubber plate and cogged wheel from the motor.
Temple rest motor
a)
Remove all the panduits
b)
Remove the PCB.
c)
Unscrew the motor axle out of the nut.
d)
Unscrew two screws from the plate
e)
Remove the motor.
f)
Remove the shaft.