CEPHALOSTAT ADJUSTMENT
7.2

Checking the sensor head position

Technical Manual
The sensor alignment tool can now be attached to the adapter pins.
The sensor head is attached to its position in reverse order:
- Push the sensor head to its adapter.
- Place the lock disc to the adapter and tighten the attachment screw with the 3mm allen key.
- Push the button axle towards the lock disk. Make sure that the head of the axle goes into the
hole on the lock disc and tighten the attachment screw with 3mm allen key.
- To remove the sensor head clearance tighten the four screws located on the lock disc
slightly with 2.5mm allen key.
- Attach the covers.
The sensor head must be perpendicular to the head support. Use a sprit level to check the
sensor head position, i.e. the sensor alignment tool position (see Fig. 121 on page E-64). In
case it is not perpendicular to the head support, the angle of the quick connector mechanism
must be adjusted.
In case the sensor head is attached to the cephalostat, remove the sensor head from the
quick connector mechanism (see instructions given in Planmeca Proline XC panoramic X-ray
unit's user's manual). Remove the covers according to the instructions given in section
"Cephalostat sensor head with quick connector mechanism - removing the covers" on page
E-58.
In case the cephalostat is equipped with fixed sensor head, remove the covers and the sen-
sor head according to the instructions given in section "Removing and attaching the fixed
sensor head" on page E-59.
Attach the sensor alignment tool to the quick connector. Loosen the two screws located on
the outer side of the mechanism with the 3mm allen key.
Figure 120
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-63