PANORAMIC X-RAY BEAM ADJUSTMENT
Chapter D - ADJUSTMENT & CALIBRATION
If you cannot adjust the beam far enough to the left using the method described above you
may have to move the metal tab on the draw nut. Loosen the securing screw rotate the tab
slightly and then tighten the screw. The collimator position should then be rechecked.
Metal tab
Positioning thread
Draw nut
Figure 26
Touch the Next field to store the collimator position in the memory and move on to the colli-
mator number 1. The number 1 will appear on the display.
Figure 27
Repeat the alignment procedure described above for the collimator number 1.
Exit the primary collimator calibration mode. Exit the service settings mode.