Checking The Soft Tissue Filter - Planmeca Proline XC Technical Manual

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

Chapter D - ADJUSTMENT & CALIBRATION

7.4 Checking the soft tissue filter

Clip the soft tissue filter positioning tool to the soft tissue filter which is at the end of the cas-
sette carriage (Fig. 78). This will enable you to see the soft tissue filter when you check its
position.
Figure 78
Touch the Prog. field on the main display. The Select program type appears. Select Ceph.
The rotating unit will move automatically to the correct position for taking cephalometric expo-
sures.
Figure 79
D-48 Planmeca Proline XC X-ray unit
Soft tissue filter
positioning tool
Soft tissue filter
CEPHALOSTAT ADJUSTMENT
Cassette carriage
Technical Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents