PATIENT POSITIONING MECHANISM
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Select a kilovolt value of 60 and a milliampere value of 2. Touch the OK field.
Figure 77
On the Test exposure display touch the OK field.
Touch OK twice to return to normal mode.
Touch the ready field to drive the X-ray unit into the exposure position. The rotating unit will
move to ready position.
Figure 78
NOTE
The ready indicator light will be flashing if the PC is not in the ready state. Refer to the
Dimaxis user's manual on how to prepare the system ready for the exposure.
Take an exposure. Hold the exposure down for the duration of the exposure.
CAUTION
Radiation is generated when exposure button is pressed.
Planmeca Proline XC X-ray unit E-37
Technical Manual