MECHANICAL ADJUSTMENTS
7.4

Vertical tube head stiffness adjustment screw

Adjust the stiffness of the vertical tube head by turning the adjustment screw on the
support axle manually or with a wrench tool.
The stiffness of the vertical tube head has been preadjusted at the factory, and can be
changed, if necessary.
Turn the adjustment screw 0,5 - 1 rounds clockwise if you want to tighten the tube head
and 0,5 - 1 rounds counterclockwise to loosen it.
NOTE Do not turn the adjustment screw too much counterclockwise to avoid the screw to
come loose.
Vertical tube head stiffness adjustment screw
Figure 11