MECHANICAL ADJUSTMENTS
7.4
Vertical tube head stiffness adjustment screw
Adjust the stiffness of the vertical tube head by turning the adjustment screw on the
support axle manually or with a wrench tool.
The stiffness of the vertical tube head has been preadjusted at the factory, and can be
changed, if necessary.
Turn the adjustment screw 0,5 - 1 rounds clockwise if you want to tighten the tube head
and 0,5 - 1 rounds counterclockwise to loosen it.
NOTE Do not turn the adjustment screw too much counterclockwise to avoid the screw to
come loose.
Vertical tube head stiffness adjustment screw
Figure 11
34
Planmeca Intra X-ray unit
Technical Manual