10 | Routine Operation
256 / 319
4
Click the Update Software button.
5
Micro GC Fusion will display the currently loaded software version.
6
Click the Choose File button and choose the software update file.
7
Click the Apply Software Update and Reboot button to complete the software
update. The instrument will reboot after the software update and may be
temporarily unavailable. Refresh the web browser 1-2 minutes after the system
reboot to reconnect with the instrument.
8
The On/Standby button on the instrument front panel will be temporarily disabled
during the system reboot. This prevents unintended instrument shutdown in the
software update process.
CAUTION
Do not shutdown or disconnect power from the instrument when a software
upgrade is in progress. Doing so may result in an incomplete upgrade and
potential data corruption.
074-594-P1H Micro GC Fusion Operating Man-
INFICON
ual