9 | Software Operation
9.10.10.7 Manual Response Factors
9.10.10.7.1 Typing a Manual Response Factor
226 / 319
Often, certain compounds are not available in a calibration gas mix. If the response
factor is known or if the response factor can be determined from another compound, a
Manual Response Factor can be used in lieu of calibration gas.
To access the Manual Response Factor:
1
Click the
icon.
2
Click Data Browser to open the data browser window.
3
Select a run to calibrate. Only one run can be calibrated at a time.
4
Click the
icon to open the calibration window.
To enter a known response factor:
1
In the calibration window, select Manual Response from the Fit Type shortcut
menu for the desired compound.
2
In the area count field, type the known response factor.
3
Click the
icon to save the method.
074-594-P1H Micro GC Fusion Operating Man-
INFICON
ual