6 | Installation
110 / 319
19
To validate that the Valco stream selector was installed successfully, open the
Micro GC Fusion Main Page on any supported web browser or wireless device.
The Valve Position quick select menu will appear, enabling the user to select a
valve position.
20
For instructions on Valco stream selector configuration, see Valco Stream
Selector Operation [} 237].
WARNING
Safely vent carrier and sample streams - potentially toxic, noxious, asphyxiant
(oxygen displacing) or flammable gases - outside Micro GC Fusion and away
from the operating area. Vent toxic gases to a fume hood, chemical trap, or
reaction medium.