9 | Software Operation
9.10.10.2.6 Start Sensitivity
208 / 319
The Start Sensitivity parameter influences the start point of peaks within a specified
retention time range. Valid Value set points are 0 to 100, with 50 being the default.
Values closer to 0 will start the peak sooner, values closer to 100 will start the peak
later. It is recommended to increment the Value by steps of 1 from the default setting
to move the start point of the peaks.
074-594-P1H Micro GC Fusion Operating Man-
INFICON
ual