6 | Installation
6.15.3 4-Module Micro GC Fusion Chassis Installation - Front
128 / 319
19
Turn on Micro GC Fusion. The instrument is ready to perform an analysis.
CAUTION
If the flow rate through the device exceeds 100 mL/min, the output pressure
from the device to Micro GC Fusion may exceed the acceptable pressure
range, causing instrument damage.
WARNING
Safely vent carrier and sample streams - potentially toxic, noxious, asphyxiant
(oxygen displacing) or flammable gases - outside Micro GC Fusion and away
from the operating area. Vent toxic gases to a fume hood, chemical trap, or
reaction medium.
Install
The pressure reducer and gas-liquid separator includes the following items. These
items are needed to install the device on the front of a 4-module Micro GC Fusion
chassis.
PN
059-0696
059-0697
059-0336
059-0335
952-229-G1
B3079191X
952-361-P5
There are three tubes in kit (PN 952-229-G1) and the shortest length tube is
used for this installation.
Tools required:
• 3/8 in. open-ended wrench
• 5/16 in. open-ended wrench
• 1/4 in. open-ended wrench
Description
5/16 in. female nut
1/16 in. ferrule set
1/16 in. Valco male nut
1/16 in. ferrule
sample tube
M3 screw
bracket E
pressure reducer and gas-
liquid separator assembly
074-594-P1H Micro GC Fusion Operating Man-
INFICON
Quantity
1
1
1
1
1
2
1
1
ual