9 | Software Operation

9.10.10.8 Auto-Calibration

9.10.10.8.1 Single Level Calibration
228 / 319
Micro GC Fusion has the ability to automatically recalibrate a previously calibrated
method. Depending on the setup of Micro GC Fusion, the auto-calibration can be set
up in different ways.
• Single level calibration
• Multi-level calibration
• Multi-level calibration using a Valco Stream Selector
1
Create a Sequence called Calibration. (See Add Sequence [} 170])
2
From the shortcut menu, select Calibration Runs.
3
Type the number of step repetitions and the method. If desired, add a delay
between each repetition.
4
Type the number of runs to average and the level in which the averaging should
take place.
ð Average Last - The number of previous runs used when averaging
component areas. This is known as "Average Replicates."
ð Level - The level in which the auto-calibration should take place. Within this
level the component areas will be replaced as long as the new values are
within 10% of the previous value. A single level calibration should have the
number 1 entered after level.
5
Click the
icon to save the sequence.
6
Run the sequence. (See Run Sequence [} 177])
074-594-P1H Micro GC Fusion Operating Man-
INFICON
ual