Inficon Micro GC Fusion 2-Module Operating Manual page 165

Gas analyzer
Table of Contents

Advertisement

INFICON
074-594-P1H Micro GC Fusion Operating Manual
Method Parameter
Temperature profile
Sample pump time
Continuous
Sample inlet heater
disabled
Sample inlet heater
temperature
Low start
temperature
Sample conditioner
enable heater
Sample conditioner
purge time
Sample conditioner
flow rate
Column Length
Maximum Ramp Rate
8 m
5°C/s
10 m
4°C/s
> 10 m
2.5°C/s
Micro GC Fusion is equipped with the Fusion Auto-Sensing Technology,
known as FAST Enabled , automatically adjusts the gain of the detector during
a sample run, providing fast and accurate analyses of gases within a large
concentration range. Prior to being FAST Enabled , the instrument provided
standard or high sensitivity settings to detect high ppm to percentage and low
ppm to high ppm components, respectively.
Unit
Description
N/A
Click on the
temperature profile.
s
Duration that the sample pump pulls sample into the
instrument
N/A
Disables sample pump and allows sample to
continuously flow through the injector, assuming
positive sample pressure
NOTE: Not available with backflush modules
N/A
Disables the sample inlet heater
°C
Temperature set point of the sample inlet
N/A
Starts the column fan to allow Micro GC Fusion to
start the run at a lower temperature
N/A
Enables the sample conditioner (if installed)
s
Duration that the sample conditioner is purged (if
installed)
N/A
Sets the sample conditioner purge flow rate to low or
high (if installed)
Software Operation | 9
icon to display the column
165 / 319

Advertisement

Table of Contents
loading