9 | Software Operation
222 / 319
8
Click the
icon to save the new calibration to the method. A window will
prompt the user to save the calibration. To override the current method, use the
original method name. An
9
Click the
icon.
10
A message will prompt the user to reload the method and apply the calibration.
11
To reprocess the calibrated run with the new information, click the
confirmation message will display Are you sure you want to reprocess the X
selected run(s)? Click the
calibration to prior runs, select the desired runs from the list before clicking the
icon.
12
A confirmation message will display.
13
Click the
icon to exit the calibration window.
14
Click the updated method preset to reload the method.
353921.34 uV
icon will indicate that there are unsaved changes.
icon to reprocess the run. To apply the new
074-594-P1H Micro GC Fusion Operating Man-
INFICON
icon. A
ual