9.4.11 Edit Sequence
074-594-P1H Micro GC Fusion Operating Manual
1
In the Methods/Sequences window, select the desired sequence.
2
Click the
icon to edit the highlighted sequence. The Sequence Editor is
displayed.
3
Click the step shortcut menu to select an action. Available actions include:
ð Load Method - Loads the desired method
ð Run Method - Runs the method selected in Load Method, once the method
set points have been reached.
ð NOTE: Separate Load Method and Run Method commands allow the user to
input a delay between the two.
ð Load and Run Method - Loads the selected method. Once the method set
points are reached, a run is initiated
ð Calibration Runs - Loads and runs the selected method. Once all runs are
complete, the area of each compound will be averaged and replaced in the
desired calibration level. See Auto-Calibration [} 228].
4
Click the method shortcut menu to select a method.
Software Operation | 9
171 / 319