9 | Software Operation
218 / 319
ð Average Replicates - Select multiple runs from the calibration gas and click
to display a window. Click
the individual compounds in the calibrated method. The runs used in average
replicates appear with a
4
The Area (μV) column will automatically populate with the new area counts for all
compounds in the run. The name of the calibration will be replaced with the
currently selected run. When using average replicates, one of the data files used
in the calculation will be used as the reference chromatogram for the calibration.
5
If using a new calibration gas standard, type the Concentration of the compound
as listed on the calibration gas standard certificate of analysis.
6
Click the
icon to save the new calibration to the method. A window will
prompt the user to save the calibration. To override the current method, use the
original method name. An
7
Click the
icon.
to confirm you want to average the areas of
icon next to each run.
icon will indicate that there are unsaved changes.
074-594-P1H Micro GC Fusion Operating Man-
INFICON
ual