9.10.10.2.7 Stop Sensitivity
074-594-P1H Micro GC Fusion Operating Manual
The Stop Sensitivity parameter influences the stopping point of peaks within a
specified retention time range. Valid Value set points are 0 to 100, with 50 being the
default. Values closer to 0 will stop the peak sooner, values closer to 100 will stop the
peak later. It is recommended to increment the Value by steps of 1 from the default
setting to move the end point of the peaks.