9.10.10.7.2 Using the Response Factor of a Known Calibrated Compound
074-594-P1H Micro GC Fusion Operating Manual
If the response factor of the desired compound is close to a previously calibrated
compound, a multiplication factor can be used to calibrate the desired compound.
To use the response factor of a calibrated compound:
1
In the calibration window, select Manual Response from the Fit Type shortcut
menu for the desired compound.
2
From the shortcut menu, select the compound to use as a reference (Ex. CO
3
In the area count field, type the known multiplication factor (Ex. 1.1).
4
Click the
icon to save to the method.
Software Operation | 9
).