Delay Field

Delay Field
Depending on the analyzer configuration, a positive or negative delay
measured in either states or time can be set. The Delay field allows you to
scroll the data and place the display window at center screen. Changing the
delay will not effect the data acquisition unless it is a timing analyzer and the
acquisition mode is automatic.
The delay range of a timing analyzer is from −2500 seconds to +2500
seconds. The delay range of a state analyzer is from −8192 states to +8192
states.
If you want to move the display window to view data located off screen to the
right, enter a positive delay. If you want to move the display window to view
data located off screen to the left, enter a negative delay.
You can enter a delay by highlighting the Delay field, then turning the knob.
You can also enter a delay by selecting the Delay field, then enter a value
from the front-panel keypad.
Delay field
Delay Field
14–9