Agilent Technologies 1660A Series User Reference page 30

50/100-mhz state, 500-mhz timing logic analyzers
Table of Contents

Advertisement

Grabbers
The grabbers have a small hook that fits around the IC pins and component
leads. The grabbers have been designed to fit on adjacent IC pins on either
through hole or surface mount components with lead spacing greater than or
equal to 0.050 inches.
Probe Cable
The probe cable contains 18 signal lines, 17 chassis ground lines and two
power lines for preprocessor use. The cables are woven together into a flat
ribbon that is 4.5 feet long. The probe cable connects the logic analyzer to
the pods, termination adapter, Agilent Technologies 10269C General-Purpose
Probe Interface, or preprocessor. Each cable is capable of carrying 0.33
amps for preprocessor power.
C A U T I O N
DO NOT exceed this 0.33 amps per cable or the cable will be damaged.
Preprocessor power is protected by a current limiting circuit. If the current
limiting circuit is activated, the fault condition must be removed. After the
fault condition is removed, the circuit will reset in one minute.
N O T E
1663A and 1664A analyzers use the same pod labels.
Minimum Signal Amplitude
Any signal line you intend to probe with the logic analyzer probes must
supply a minimum voltage swing of 500 mV to the probe tip. If you measure
signal lines with a voltage swing of less than 500 mV, you may not obtain a
reliable measurement.
Maximum Probe Input Voltage
The maximum input voltage of each logic analyzer probe is ±40 volts peak.
Pod Thresholds
Logic analyzer pods have two preset thresholds and a user-definable pod
threshold. The two preset thresholds are ECL (−1.3 V) and TTL (+1.5 V).
The user-definable threshold can be set anywhere between −6.0 volts
and +6.0 volts in 0.05 volt increments.
All pod thresholds are set independently.
General-Purpose Probing System Description
Probing
2–7

Advertisement

Table of Contents
loading

This manual is also suitable for:

1660a1661a1662a1663a1664a1663as

Table of Contents