Agilent Technologies 1660A Series User Reference page 372

50/100-mhz state, 500-mhz timing logic analyzers
Table of Contents

Advertisement

Specifications and Characteristics
Specifications and Characteristics
Timers Qty. 2. Timers may be Started, Paused, or Continued at entry
into any sequence level after the first.
Timer Range 400 ns to 500 s
Timer Resolution 16 ns or 0.1% whichever is greater
Timer Accuracy +/- 32 ns or +/- 0.1%, whichever is greater
Timer Recovery Time 70 ns
Data In to Trigger Out BNC Port 110 ns typical
Measurement and Display Functions
Arming Each analyzer can be armed by the Run key, the other analyzer,
or the Port In.
Trace Mode Single mode acquires data once per trace specification;
repetitive mode repeats single mode acquisitions until stop is pressed or
until pattern time interval or compare stop criteria are met.
Labels Channels may be grouped together and given a 6-character
name. Up to 126 labels in each analyzer may be assigned with up to 32
channels per label. Trigger terms may be given an 8-character name
Activity Indicators Provided in the Configuration, State Format, and
Timing Format menus for monitoring device-under-test activity while
setting up the analyzer.
Markers Two markers (X and O) are shown as dashed lines in the
display.
Trigger Displayed as a vertical dashed line in the timing waveform,
state waveform and X-Y chart displays and as line 0 in the state listing
and state compare displays.
19–10