To run the self-tests (1664A)

Self-tests identify the correct operation of major functional areas of the
instrument. You can run all self-tests without accessing the interior of the
instrument. If a self-test fails, the troubleshooting flowcharts instruct you to
return the instrument to a qualified Agilent Technologies service center.
If you just did the power-up self-tests, go to step 2.
1
If you did not just do the power-up self-tests, disconnect all inputs,
insert the operating system disk, then turn on the power switch. Wait
until the power-up tests are complete.
Press the System key, then select the field next to System. Select
2
Test in the pop-up menu and then press the Select key.
Select the box labeled Load Test System then press the select key.
3
Insert the disk containing the performance verification tests
(self-tests) into the disk drive (normally the same as the boot disk).
Select the box labeled Continue and press the Select key. After the
4
test files have been loaded, and Analy menu is displayed.
You can run all the analyzer tests at one time by selecting All Analyzer Tests.
To see more details about each test, you can run each test individually.
Operator's Service
To run the self-tests (1664A)
20-19