Pod Threshold Field

Pod Threshold Field
The pod threshold field is used to set a voltage level that the data must reach
before the analyzer recognizes and displays it as a change in logic levels. You
specify a threshold level for each pod in a pod pair. The level specified for
each pod is also assigned to the pod's clock threshold.
When the Pod Threshold field is selected, a threshold selection pop-up
appears with the following choices:
TTL
When TTL is selected as the threshold level, the data signals must reach +1.5
volts.
ECL
When ECL is selected as the threshold level, the data signals must reach
−1.3 volts.