The Format Menu
Setup/Hold Field (State only)
Setup/Hold Field (State only)
Setup/Hold adjusts the relative position of the clock edge with respect to the
time period that data is valid. When the Setup/Hold field is selected, a
configuration menu appears. Use this Setup/Hold configuration menu to
select each pod in the analyzer and assign a Setup/Hold selection from the
selection list.
With a single clock edge assigned, the choices range from
3.5-ns Setup/0.0-ns Hold to 0.0-ns Setup/3.5-ns Hold. With both edges of a
single clock assigned, the choices are from 4.0-ns Setup/0.0-ns Hold to 0.0-ns
Setup/4.0-ns Hold. If the analyzer has multiple clock edges assigned, the
choices range from 4.5-ns Setup/0.0-ns Hold to 0.0-ns Setup/4.5-ns Hold.
Setup and hold menu
Setup / Hold field
Setup and Hold Menu
11–23