Operator's Service
To run the self-tests (1660A through 1663A)
Select Board Tests, then select Run. When the Board Tests are
13
finished, select Done.
Select Data Input Inspection. All lines should show activity. Select
14
Done to exit the Data Input Inspection.
To exit the tests, press the System key. Select the field to the right of
18
the Sys PV field.
Remove the disk containing the performance verification test, then
19
select the Exit Test System.
If you are performing the self-tests as part of the troubleshooting flowchart,
return to the flowchart.
20-18