The Listing Menu

Markers Field

Markers Field
The Markers field is used to access the markers selection menu. When the
Markers field is selected, a marker selection menu appears with the marker
choices available with the present analyzer configuration.
State Analyzer Types
In a state analyzer with count Time and count State turned off in the Trigger
menu, only Pattern markers are available. With count Time turned on,
additional choices of Time markers and Statistics markers become available.
With count States turned on, in addition to Pattern markers there are States
markers available.
Timing Analyzer Types
If a timing analyzer you have marker choices of Pattern, Time, or Statistics.
Off
The Off selection turns off marker operations. If a stop measurement was
previously specified, and the stop measurement criteria are met, the
measurement will stop even though the markers are off.
Markers Field
Markers selection menu
Markers Field
13–4