The Trigger Menu
Arming Control Field
Arming Control Using External BNCs
A more complex arming example involves passing arm signals in and out
through the External BNCs on the rear panel.
The first analyzer is armed by an Arm In signal from an external trigger
source or another test/measurement entity. After the first analyzer triggers,
it arms the second analyzer. After the second analyzer triggers, it can send a
Port Out signal to an external BNC which is used to arm another external
test/measurement entity.
The Arm Out signal is generated by either one of the two local analyzers
which you select.
After you configure the Arming Control menu, a graphical representation of
the analyzer arming configuration is displayed as shown below.
Port In/Out Arming
12–35