To run the self-tests (1660A through 1663A)
Self-tests identify the correct operation of major functional areas of the
instrument. You can run all self-tests without accessing the interior of the
instrument. If a self-test fails, the troubleshooting flowcharts instruct you to
return the instrument to a qualified Agilent Technologies service center.
If you just did the power-up self-tests, go to step 2.
1
If you did not just do the power-up self-tests, disconnect all inputs,
then turn on the power switch. Wait until the power-up tests are
complete.
Press the System key, then select the field next to System. Then,
2
select Test in the pop-up menu.
Operator's Service
To run the self-tests (1660A through 1663A)
20-13