Agilent Technologies 1660A Series User Reference page 390

50/100-mhz state, 500-mhz timing logic analyzers
Table of Contents

Advertisement

Operator's Service
To run the self-tests (1660A through 1663A)
Insert the disk containing the performance verification tests
3
(self-tests) into the disk drive. Select the box labeled Load Test
System, then select Continue.
Press the System key, then select then field next to Sys PV. Select
4
System Test to access the system tests.
Select ROM Test. The ROM Test screen is displayed.
5
You can run all tests at one time by running All System Tests. To see more
details about each test, you can run each test individually. This example
shows how to run an individual test.
20-14