Check the setup/hold combination

Select the logic analyzer setup/hold time.
1
a In the logic analyzer Format menu, select Master Clock.
b Select the Setup/Hold field, then select the setup/hold combination to be tested for all
pods. The first time through this test, use the top combination in the following table.
Setup/Hold Combinations
3.5/0.0 ns
0.0/3.5 ns
1.5/2.0 ns
c Select Done to exit the setup/hold combinations.
Disable the pulse generator channel 2 COMP (with the LED off).
2
Using the Delay mode of the pulse generator channel 1, position the pulses
3
according to the setup time of the setup/hold combination selected, +0.0 ps or −100
ps.
a In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set
Marker 1 at −1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at
−1.3000 V.
b In the oscilloscope Delta T menu, select Start on Pos Edge 1. Select Stop on Pos
Edge 1.
To test the single-clock, single-edge, state acquisition
Testing Performance