6.2

Before the use

6.2.1 Assembling of O2 cell

Insert and screw the cell: in the space
marked with the script "O 2 CELL".
Connect the electric cable of the cell passing
it into dedicated space.
The O 2 cell connector
The RJ connector on the specific socket
within the valves group.
WARNING! Risk of injury for the patient
Check of electric connection of O 2 cell.
At each anaesthesia unit start-up, the system checks the presence of
the electric connection to the O 2 cell ( "SELF TEST" phase : cfr. 6.4.1
or cfr. 7 ).

6.2.2 Assembling of absorber canister

On the anaesthesia unit it is possible to use a multipurpose
rechargeable absorber canister or a disposable pre-loaded absorber
canister.
It is recommended to use only original SIARE spare parts or spare parts
checked and approved by SIARE.
The absorber canister is positioned in the upper side of valves group. The specific
lock (unlock) lever makes very easy to hook (unhook) the canister.
Mount the CO 2 absorber canister on the
anaesthesia unit as shown in the picture.
To unpack carefully the CO2 absorber
canister.
Shake the absorber canister (disposable
model) in order to separate the soda lime
granules.
Remove the seal from absorber canister
(disposable model).
Insert the absorber canister in the apposite
groove (opposite side of the "PUSH" lever)
6-4