Turbine Over-
temperature
Turbine Over-
current
Low Gas Supply
Can-Bus Failure
MORPHEUS
This is a high priority alarm which is activated after a 0 sec
delay starting from the max limit of the turbine temperature
(105°C) and it is restored once the temperature goes under the
100°C.
It is indicated by an acoustic and visual alarm and it can be
silenced but not inhibited.
This is a high priority alarm which is activated after a 700 ms
delay starting from the max limit of the voltage during the
turbine operation.
It is indicated by an acoustic and visual alarm and it can be
silenced but not inhibited.
This high priority alarm occurs in case the supply pressure of
the gas channel remains for more than 5 sec. (delay time) at a
lower value than the minimum necessary one to allow the
equipment operation.
It is indicated by an acoustic and visual alarm. It can be silenced
(suspended) for 120 sec. If the FiO2 % is set on 21% such
alarm will be automatically inhibited.
This high priority alarm occurs in case when the CAN system
doesn't recognise the correct adjustments in the electronic
boards.
WARNING! Risk of injury for the patient
An auxiliary system of ventilation must be always available for the
patient when the ventilator is for a total life support.
WARNING! Risk of injury for the unit
Switch off the device immediately and get in
touch with the technical service urgently.
WARNING! Risk of injury for the unit
Switch off immediately the device and get in
touch with the technical service urgently.